Z-1925-2017: Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

Recall record Z-1925-2017 for Magellan Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1925-2017
Company in FDA record
Magellan Diagnostics, Inc.
FDA classification date
May 18, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Source facts

Action
Magellan issued a follow-on Recall notification related to the FDA Safety Notification dated May 17, 2017. This current recall follows FDAs recommendation to remove venous blood samples as an allowable sample type from all LeadCare products. This recall will supersede and replace the prior Recall notifications. Notifications issued by : UPS Letter (US), Registered Letter, E-mail, or Phone calls form May 23-May 25. Questions: Reba Daoust, Director of QA/RA 978-248-4811 or rdaoust@magellandx.com Refer to Firm's website: http://www.leadcare2.com/getmedia/2fd8a90f-5c64-4058-b6ab-5db13512f0fe/Letter-to-our-customers-5-17-17.pdf.aspx. The notification of May 2017 supercedes the firms previous communications:. Magellan Diagnostics, notified customers by letter dated Nov 29, 2014 and provided with the problem, mitigation action( customers have been advised to implement a minimum 24-hour incubation of the blood-treatment reagent mixture prior to analysis) . Customers were instructed to return a completed FaxBack form via fax to 978-600-1480. A revised letter issued May 1, 2017 recommending that customers make the following workflow changes: 1. Discontinue collecting blood samples for lead testing in Becton Dickinson (BD) EDTA Vacutainer tubes: Lavender and Tan tops. 2. Discontinue the 24 hour incubation for sample preparation. On 5/17/17 FDA issued News Release - FDA warns Americans about risk of inaccurate results from certain lead tests- manufactured by Magellan Diagnostics may provide inaccurate results for some children and adults. Blood drawn from a vein may provide results that are lower than the actual level in the blood. FDA is warning laboratories not to use Magellan Diagnostic lead tests wit blood drawn from the vein. At this time all Leadcare systems can be used with blood drawn from a finger or heel stick. Based on FDA/s warning the firm issued notification letter to users on 5/xx/17.
Initiated
2014-11-29
Root cause
Device Design
Terminated
2022-02-26
Distribution
Nationwide Distribution
Recall number
Z-1925-2017
Recall status
Terminated
CDRH record ID
154782
FDA classified
2017-05-18
Recall event number
76991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗