Skip to content

Public FDA record

Z-1925-2021: reSET Mobile App, iOS and Android, provides cognitive behavioral therapy, as part of substance abuse trea

Recall record Z-1925-2021 for Pear Therapeutics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1925-2021
Company in FDA record
Pear Therapeutics, Inc.
FDA record creation date
Jun 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, patients with a urine drug screen (UDS) received access to a contingency management wheel spin for potential rewards regardless of the UDS results. Positive UDS results should not have resulted in access to a wheel spin.

Source facts

Action
On 05/20/2021, the Recalling Firm communicated via email an initial "URGENT: MEDICAL DEVICE RECALL" Notification with an amended/follow-up Notification Letter sent on or about 06/07/2021 informing customers that between 08/10/2020 and 04/30/2021, patients with a urine drug screen (UDS) logged into the Recalling Firm's database received access to a contingency management wheel spin for potential rewards regardless of the UDS results. Positive UDS results should not have resulted in access to a wheel spin. Customers are also being informed that there are no actions required for the safe handling of the corrected product. A software release was issued on 04/30/2021 to return the products to their intended operation. Given the potential for reduced clinical effectiveness, Healthcare Providers are encouraged to review patients that have a positive urine drug screen logged to ensure that the patient receives any treatment or counseling necessary to support the patient's treatment. Customers are asked to complete and return the "REPLY FORM" via Fax at 415-458-2700 (Attn: Yuri Maricich) or email to postmarked@peartherapeutics.com. For questions or further assistance, call 1-833-697-3738 (available weekdays 8:00 a.m. - 8:00 p.m. EST)
Initiated
2021-06-07
Root cause
Software design
Terminated
2023-06-14
Distribution
U.S.: AZ, CA, DE, FL, IL, IN, KY, MA, MD, MN, NC, NY, OH, TN, WA, and WI O.U.S.: N/A
Recall number
Z-1925-2021
Recall status
Terminated
CDRH record ID
187802
FDA record created
2021-06-21
Recall event number
88061

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗