Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.
Source facts
Action
Spine Frontier sent an email dated December 10, 2012, to the affected customers that a Pedfuse Inserter with part number PM3979 needed to be returned.
Further questions please call (978) 232-3990.
Initiated
2013-03-14
Root cause
Device Design
Terminated
2015-06-03
Distribution
Distribution only in Texas.
Recall number
Z-1926-2013
Recall status
Terminated
CDRH record ID
117582
FDA classified
2013-08-08
Recall event number
64984
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.