Open original FDA record ↗Recall record
- FDA identifier
- Z-1926-2019
- Company in FDA record
- Infinium Medical, Inc
- FDA record creation date
- Jul 2, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The CLEO & Tranquility VS Patient Monitors 510(k) indications for use in the United States did not include the temperature monitor accessory option.
Source facts
- Action
- Infinium Medical, Inc. notified customers on about 05/31/2019 via "URGENT: MEDICAL DEVICE RECALL" letter sent through email and certified letter. Instructions included to remove all temperature modules from the CLEO & Tranquility VS Patient Monitors and destroyed. They were also requested to complete and return the Medical Device Recall Return Response Form to QARA@infiniummedical.com. Once the response form is received, a new IFU and standalone temperature monitor will be provided. Questions and comments can be sent to Mr. George Anello during regular business hours (8:00 AM - 04:30 PM EST) at QARAinfiniummedical.com or by phone at l-866-918-8434 ext. 109.
- Initiated
- 2019-05-31
- Root cause
- Labeling False and Misleading
- Distribution
- Distributed nationwide to California, Georgia, Louisiana, Florida, Maryland, Minnesota, Puerto Rico, South Carolina, Virginia, Wisconsin.
- Recall number
- Z-1926-2019
- Recall status
- Open, Classified
- CDRH record ID
- 173303
- FDA record created
- 2019-07-02
- Recall event number
- 82999