Skip to content

Public FDA record

Z-1926-2025: DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20

Recall record Z-1926-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1926-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jun 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.

Source facts

Action
An "URGENT MEDICAL DEVICE RECALL (UMDR)" notification letter dated 4/16/25 was sent to customers. ACTION: If your laboratory encounters the system event described in this letter: Perform a system initialization to resume normal operation. Repeat testing for any canceled tests. No action is required if your laboratory has not observed the system event listed above. RESOLUTION: Beckman Coulter will release a modification for this issue. Your service representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact Customer Support Center: From our website: http://www.beckmancoulter.com
Initiated
2025-04-23
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries of: Italy, Germany, Spain, New Zealand, France, Australia, Slovakia, Switzerland, Ireland, Brazil, Taiwan, Province of China, India, United Kingdom of Great Britain and Northern Ireland, Hungary, Austria, Portugal, Croatia, Kazakhstan, Cote d'lvoire, Poland, Korea, Republic of, Czechia, Algeria, Libya, Oman, Singapore, and Belgium.
Recall number
Z-1926-2025
Recall status
Open, Classified
CDRH record ID
213550
FDA classified
2025-06-06
Recall event number
96760

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗