Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.
Source facts
Action
Maquet/Getinge notified customers via an Urgent Medical Device Recall Notice dated November 16, 2017.
Initiated
2017-11-16
Root cause
Unknown/Undetermined by firm
Terminated
2021-07-21
Distribution
U.S. and Gov. accts.
Recall number
Z-1927-2018