Open original FDA record ↗Recall record
- FDA identifier
- Z-1927-2019
- Company in FDA record
- Roche Diagnostics Corporation
- FDA classification date
- Jul 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Source facts
- Action
- On April 4, 2019, Roche issued Urgent Medical Device Correction notices (UMDC TP-00620) to customers via UPS Ground (receipt required). Customers were advised to take the following actions: 1) Inspect the tray input flap on the cobas p 501 or p 701 post-analytical unit to determine if it is loose. If it is not fastened securely or becomes loose prior to the part being replaced, contact your Roche Field Engineer Specialist (FES) to repair the instrument. 2) Complete and return the fax form (TP-00621) via fax or email 3) File the UMDC for future reference. Customers with questions may contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in the UMDC.
- Initiated
- 2019-04-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-05-12
- Distribution
- US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
- Recall number
- Z-1927-2019
- Recall status
- Terminated
- CDRH record ID
- 172194
- FDA classified
- 2019-07-03
- Recall event number
- 82651