Open original FDA record ↗Recall record
- FDA identifier
- Z-1927-2021
- Company in FDA record
- GE Healthcare, LLC
- FDA record creation date
- Jun 22, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a certain sequence of events occur, the pdf test report that belongs to one patient will appear in a different patient's record when viewed in an Electronic Medical Record (EMR) or a similar system.
Source facts
- Action
- The recalling firm issued letters dated 4/19/2021 beginning 4/19/2021. For systems that are on Windows 8.1 and above, the issue will be corrected through a software patch. For the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7, the letter instructs that the devices are not to be networked on these systems and that a software patch cannot be made available to these systems since the operating systems are not supported by Microsoft. A second cover letter and customer notification were issued dated 5/10/2021 updating the second safety instruction regarding the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7 saying that these systems must be rebooted before manually exporting tests.
- Initiated
- 2021-04-19
- Root cause
- Software design
- Terminated
- 2026-01-14
- Distribution
- Worldwide distribution - US Nationwide. Government/military distribution was also made, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Brazil, Brunei Darussalam, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Czechia, Denmark, Estonia, Fiji, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Mexico, Moldova, Myanmar, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam.
- Recall number
- Z-1927-2021
- Recall status
- Terminated
- CDRH record ID
- 187234
- FDA record created
- 2021-06-22
- Recall event number
- 87844