Open original FDA record ↗Recall record
- FDA identifier
- Z-1928-2009
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA classification date
- Aug 27, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software bug: A unique series of events involving the COBAS AmpliPrep instrument running AMPLILINK software version 3.2.2 that results in specimens being amplified and detected with the wrong parameters. The error can occur on either COBAS AmpliPrep/COBAS TaqMan 48, COBAS AmpliPrep/COBAS TaqMan (docked) or cobas s 201 system (docked) running Amplilink Software versions 3.1 or 3.2 series up to and
Source facts
- Action
- Urgent Medical Device Correction Letters were sent on June 2, 2009 by first class mail. The letter described the issue; impact on results; clinical implications; the workaround; and the actions required, which include following the workaround until a new version of AMPLILINK software is available and filing the letter for future reference. Questions or concerns should be directed to Roche Diagnostics Technical Support at 1-800-526-1247.
- Initiated
- 2009-06-02
- Root cause
- Software design
- Terminated
- 2010-08-13
- Distribution
- Nationwide Distribution
- Recall number
- Z-1928-2009
- Recall status
- Terminated
- CDRH record ID
- 82140
- FDA classified
- 2009-08-27
- Recall event number
- 52118