Reason stated in the source
The adaptive volume/pressure regulation can be automatically and permanently reduced to PEEP + 3 mbar when a closed suctioning maneuver is performed on the patient and may be undetected by the operator.
Source facts
- Action
- A Medical device Safety Alert and Corrective Action along with a cover letter, was sent to all locations on February 16, 2011. The letter identified the affected product along with the reason for recall. The letter states that operators deactivate the "high tidal volume" alarm as described in appendix K.4 of the operator's manual until the ventilator is upgraded with a corrected software. Distributors will upgrade the affected devices with the corrected software as soon as possible. Questions can be directed to 800-426-6331, ext 215.
- Initiated
- 2011-02-16
- Root cause
- Software design
- Terminated
- 2012-03-21
- Distribution
- Nationwide Distribution -- MA, FL, GA, NC, MI, NC, MO, TX, and WV.
- Recall number
- Z-1928-2011
- Recall status
- Terminated
- CDRH record ID
- 97973
- FDA classified
- 2011-04-08
- Recall event number
- 58014