Open original FDA record ↗Recall record
- FDA identifier
- Z-1928-2016
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Jun 8, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.
Source facts
- Action
- GE Healthcare sent an " Urgent Medical Device Correction" letter dated May 9, 2016, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Detail, Product Correction & Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2016-05-09
- Root cause
- Nonconforming Material/Component
- Terminated
- 2018-07-11
- Distribution
- US: AL, AZ, CA, CO, FL, ID, IL, IA,ME, NE, NV,NH, NJ, NY, NC, OH, PA, SC, SD, TN, TX, UT,VA, WA, WI. OUS: Argentina Australia Austria Brazil Canada Chile Colombia Denmark Egypt Finland France Germany Greece Hong Kong Iceland Italy Japan Korea (Republic of) Kuwait Malaysia Netherlands Norway Peru Poland Portugal Russia Singapore South Africa Spain Sweden Switzerland Ukraine United Arab Emirates United Kingdom
- Recall number
- Z-1928-2016
- Recall status
- Terminated
- CDRH record ID
- 146825
- FDA classified
- 2016-06-08
- Recall event number
- 74215