Skip to content

Public FDA record

Z-1928-2019: cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.

Recall record Z-1928-2019 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1928-2019
Company in FDA record
Roche Diagnostics Corporation
FDA classification date
Jul 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.

Source facts

Action
On April 4, 2019, Roche issued Urgent Medical Device Correction notices (UMDC TP-00620) to customers via UPS Ground (receipt required). Customers were advised to take the following actions: 1) Inspect the tray input flap on the cobas p 501 or p 701 post-analytical unit to determine if it is loose. If it is not fastened securely or becomes loose prior to the part being replaced, contact your Roche Field Engineer Specialist (FES) to repair the instrument. 2) Complete and return the fax form (TP-00621) via fax or email 3) File the UMDC for future reference. Customers with questions may contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in the UMDC.
Initiated
2019-04-04
Root cause
Nonconforming Material/Component
Terminated
2020-05-12
Distribution
US Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
Recall number
Z-1928-2019
Recall status
Terminated
CDRH record ID
172195
FDA classified
2019-07-03
Recall event number
82651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗