Z-1928-2021: Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent

Recall record Z-1928-2021 for Quidel Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1928-2021
Company in FDA record
Quidel Corporation
FDA classification date
Jun 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used. These specimens would be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5. There have been no formal complaints of False Negative results with Ct values <5 from Lyra SARS-CoV-2 Assay users.

Source facts

Action
On May 13, 2021, Quidel issued an Urgent Medical Device Correction notice to customers via email notifying them that Quidel has determined that there is a potential for patients with high SARS CoV 2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5 when using specified applications. YOU NEED TO TAKE THE FOLLOWING ACTIONS: 1. Immediately review the revised Lyra SARS CoV 2 Assay Instructions for Use (IFU) for updated Result Interpretation guidance: https://www.fda.gov/media/136820/download. 2. For clinical specimens generating Ct values d5, perform 1:10 and 1:100 dilutions using uninoculated transport media, extract, and test the diluted samples according to the Lyra SARS CoV 2 IFU. Refer to the Lyra SARS CoV 2 Assay IFU for full, detailed instructions. 3. It is recommended that any data which has been previously generated be re-evaluated and samples re-tested, as needed. 4. Complete the enclosed Field Corrective Action Fax Back form acknowledging receipt and understanding of the recommended actions. Return to Quidel Technical Support via fax at 858.203.9297 or via email to customernotifications@quidel.com.
Initiated
2021-04-27
Root cause
Device Design
Terminated
2022-09-08
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, CT, DC, FL, ID, IN, MA, MI, MO, NM, NY, OH, SC, TN, TX, WA and the countries of Canada, Saudi Arabia, United Kingdom.
Recall number
Z-1928-2021
Recall status
Terminated
CDRH record ID
187617
FDA classified
2021-06-26
Recall event number
87996

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗