Z-1929-2016: ExacTrac Vero is a Patient Positioning System for Radiation therapy.

Recall record Z-1929-2016 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1929-2016
Company in FDA record
Brainlab AG
FDA record creation date
Jun 8, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potentially incorrect positioning when using Implanted Marker Detection with Brainlab ExacTrac Vero 3.5

Source facts

Action
A FIELD SAFETY NOTICE / PRODUCT NOTIFICATION was sent to customers on May 4, 2016 and distributed via email. An active reply by the consignees will be followed, if necessary via further different communication media. As of now for new installations - the Product Notification letter is handed over to the customer during installation (before clinical use) by the according Brainlab Service representative, - a signature is required by the customer that the contained User Corrective Action will be implemented before clinical use by the facility (to be adhered to by the users). Brainlab Instructions and actions required to be performed by the customer are described below. 1. For Positioning and Verification, carefully review plausibility of implant marker fusions resulting in rotation angles of more than 10. If in doubt consider the following options before proceeding: - Deselect one or several markers defined in the CT data set, if you suspect these markers to have migrated - Apply bony fusion instead of marker based fusion - Apply ConeBeam CT correction and verification instead of marker based fusion - Acquire a new CT scan and repeat treatment planning and definition of implanted marker positions based on the new CT data. For further details regarding the suggested proceedings, please consult the according Clinical User Guide. 2. For Dynamic Tracking continue to follow the instructions of the Clinical User Guide regarding the review of the correlation model based on implanted marker detections. Specifically: - Examine the correctness of localization of markers and targets in the X-ray sequence. - Verify the consistency of the Target Detection Curve and/or Correlation Model Curve. Irregularities of the curves are an indication for incorrect implanted marker detection. In this case please consider the options described in 1. before proceeding. If the customers require further clarification, they should contact their local Brainlab Customer
Initiated
2016-05-04
Root cause
Device Design
Terminated
2017-07-24
Distribution
Distributed in the states of Florida, New York, Texas and Ohio, and in the countries of Belgium, France, Germany, Italy, Japan and South Korea.
Recall number
Z-1929-2016
Recall status
Terminated
CDRH record ID
147016
FDA record created
2016-06-08
Recall event number
74265

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗