Skip to content

Public FDA record

Z-1929-2017: Olympus ENDOEYE HD II Video Telescopes Models: WA50040A Endoeye HD II, 10 mm, 0 deg., autoclavable; WA500

Recall record Z-1929-2017 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1929-2017
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Feb 21, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A damaged temperature sensor at the endoscope's tip can cause the endoscope's distal end to become abnormally hot.

Source facts

Action
A letter to consignees was sent February 10, 2017. Olympus will be repairing the device at user sites by disabling the temperature sensor.
Initiated
2017-02-10
Root cause
Device Design
Terminated
2018-05-09
Distribution
Nationwide
Recall number
Z-1929-2017
Recall status
Terminated
CDRH record ID
153345
FDA classified
2017-02-21
Recall event number
76488

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗