Reason stated in the source

Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta.

Source facts

Action
Customers were notified on about 03/30/2018 via an Urgent Medical Device Safety Notice. Instructions included to review the field safety notice and understand the potential hazard, review the recall with appropriate clinical staff at their institution, complete and return the acknowledgement form, and distribute the notification within the organization where potentially affected devices may have been transferred. Questions can be directed to Edwards Customer Service, Monday through Friday at 800-424-3278 form 8:00 AM - 4:00 PM Pacific Time.
Initiated
2018-03-30
Root cause
Under Investigation by firm
Terminated
2023-11-29
Distribution
Worldwide Distribution -- USA, to the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MI, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV; and, the countries of to Australia, Austria, Belgium, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Recall number
Z-1929-2018
Recall status
Terminated
CDRH record ID
162896
FDA record created
2018-05-24
Recall event number
79650

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗