Open original FDA record ↗Recall record
- FDA identifier
- Z-1929-2018
- Company in FDA record
- Edwards Lifesciences, LLC
- FDA record creation date
- May 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta.
Source facts
- Action
- Customers were notified on about 03/30/2018 via an Urgent Medical Device Safety Notice. Instructions included to review the field safety notice and understand the potential hazard, review the recall with appropriate clinical staff at their institution, complete and return the acknowledgement form, and distribute the notification within the organization where potentially affected devices may have been transferred. Questions can be directed to Edwards Customer Service, Monday through Friday at 800-424-3278 form 8:00 AM - 4:00 PM Pacific Time.
- Initiated
- 2018-03-30
- Root cause
- Under Investigation by firm
- Terminated
- 2023-11-29
- Distribution
- Worldwide Distribution -- USA, to the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MI, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV; and, the countries of to Australia, Austria, Belgium, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Recall number
- Z-1929-2018
- Recall status
- Terminated
- CDRH record ID
- 162896
- FDA record created
- 2018-05-24
- Recall event number
- 79650