Z-1929-2019: Vereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tom

Recall record Z-1929-2019 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1929-2019
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA record creation date
Jul 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.

Source facts

Action
On June 3, 2019, the firm notified their customers of the recall via an "Urgent Field Safety Notice Medical Device Correction" letter. Customers were advised to take the following actions: * The operator is instructed to observe the patient during motorized motion. In case of collision, the operator should use the E-Stop to abort all motions. " The operator is instructed to ensure accessories do not collide with gantry covers. Additionally, ensure that sheets, blankets and/or IV lines are not loose or hanging outside the Patient Pallet. " Ensure the patients arms/hands are positioned appropriately at their side or overhead. If necessary, use the arm restraint strap as described in the IFU. " Instruct the patient to not move their hands during the scan and during any table motions until the operator removes the patient from the scanner at the completion of the study. Philips will deploy Field Change Order (FCO) 88200520 free of charge to address overall cover fit on the affected systems. A Philips Field Service Engineer will contact you to schedule the updates at your site. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
Initiated
2019-06-03
Root cause
Device Design
Terminated
2021-04-26
Distribution
Worldwide Distribution - US Nationwide Domestic distribution to CA GA LA MD MI MN MO NE NJ NY OH TN TX VT WI. Foreign distribution to Mexico, Argentina Austria Belgium China Colombia France Germany Israel Italy Japan Latvia Luxembourg Netherlands Portugal Singapore Spain Switzerland
Recall number
Z-1929-2019
Recall status
Terminated
CDRH record ID
173945
FDA record created
2019-07-03
Recall event number
83187

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗