Z-1930-2019: RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (Ra

Recall record Z-1930-2019 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1930-2019
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Jul 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong

Source facts

Action
RaySearch issued Field Safety Notice, Medical Device Correction #24138 via email on November 20th, 2017. The letter identifies the product, issue, health risk and action to take: Do not use a number of histories above 6 million for electron dose calculation. Inspect all commissioned electron beam models for any applicator calculated with greater than 6 million. Contact customer support if you have any further questions. Confirm you have read and understood this notice. This issue will be resolved in the next version of RayStation, scheduled for market release in December 2017. Educate planning staff and all users about this workaround.
Initiated
2017-11-20
Root cause
Software design
Terminated
2023-08-08
Distribution
US Nationwide Distribution
Recall number
Z-1930-2019
Recall status
Terminated
CDRH record ID
173846
FDA record created
2019-07-03
Recall event number
83152

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗