Open original FDA record ↗Recall record
- FDA identifier
- Z-1930-2019
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA record creation date
- Jul 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
Source facts
- Action
- RaySearch issued Field Safety Notice, Medical Device Correction #24138 via email on November 20th, 2017. The letter identifies the product, issue, health risk and action to take: Do not use a number of histories above 6 million for electron dose calculation. Inspect all commissioned electron beam models for any applicator calculated with greater than 6 million. Contact customer support if you have any further questions. Confirm you have read and understood this notice. This issue will be resolved in the next version of RayStation, scheduled for market release in December 2017. Educate planning staff and all users about this workaround.
- Initiated
- 2017-11-20
- Root cause
- Software design
- Terminated
- 2023-08-08
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1930-2019
- Recall status
- Terminated
- CDRH record ID
- 173846
- FDA record created
- 2019-07-03
- Recall event number
- 83152