Open original FDA record ↗Recall record
- FDA identifier
- Z-1931-2024
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- May 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Source facts
- Action
- On May 2, 2024, CareFusion, a subsidiary of Becton, Dickinson and company issued a "Urgent Medical Device Correction", to affected consignees. On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.
- Initiated
- 2024-04-15
- Root cause
- Component design/selection
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of South Africa, Kuwait, Qatar, Singapore, Bermuda, United Arab Emirates, Argentina, Australia, Belgium, Bahrain, Brazil, Bahamas, Canada, Chile, China, Germany, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Italy, Japan, South Korea, Lebanon, Morocco, Mexico, New Zealand, Oman, Portugal, Thailand, Taiwan.
- Recall number
- Z-1931-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207442
- FDA classified
- 2024-05-30
- Recall event number
- 94527