Z-1932-2019: Medtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the s

Recall record Z-1932-2019 for Medtronic Navigation, Inc.-Littleton. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1932-2019
Company in FDA record
Medtronic Navigation, Inc.-Littleton
FDA record creation date
Jul 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced

Source facts

Action
Medtronic Navigation issued letter May l, 2019 via FedEx Express Overnight delivery identifying the issue, health risk and action to take: provided a visual mitigation card to attach to the Image Acquisition System (IAS) of their affected systems to serve as a visual reminder. " Consignees will be asked to sign and return the consignee response form; acknowledging receipt of this information and confirmation of having attached the visual mitigation card to their affected system. Review and retain Attachment 8 for further guidance . Attachment B provides instructions for inspecting the battery and chargers. and follow the "Performance Checks and Maintenance" section of your user manual, in addition to the information in Attachment B, to ensure your system is available for surgical procedures Medtronic field service personnel will install a new version of the user manual " Medtronic field service personnel will install a component (Mobile View Station \'MVS'') heatsink) as part of the new design of the AC power input circuit. Questions contact Medtronic Technical Support if you have any questions or concerns regarding this matter: USA: (800) 595-9709 or 720-890-3160 Worldwide: +1-720-890-3160 RS.NA VTECHSUPPORT@M EDTRON IC.COM
Initiated
2019-05-01
Root cause
Device Design
Terminated
2021-01-19
Distribution
Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KOREA KOREA, REPUBLIC OF KUWAIT LEBANON LIBYAN ARAB JAMAHIRIYA LK MEXICO NETHERLANDS NEW ZEALAND NICARAGUA POLAND PORTUGAL PUERTO RICO RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVAKIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM VENEZUELA
Recall number
Z-1932-2019
Recall status
Terminated
CDRH record ID
173294
FDA record created
2019-07-03
Recall event number
82991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗