Open original FDA record ↗Recall record
- FDA identifier
- Z-1932-2019
- Company in FDA record
- Medtronic Navigation, Inc.-Littleton
- FDA record creation date
- Jul 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
Source facts
- Action
- Medtronic Navigation issued letter May l, 2019 via FedEx Express Overnight delivery identifying the issue, health risk and action to take: provided a visual mitigation card to attach to the Image Acquisition System (IAS) of their affected systems to serve as a visual reminder. " Consignees will be asked to sign and return the consignee response form; acknowledging receipt of this information and confirmation of having attached the visual mitigation card to their affected system. Review and retain Attachment 8 for further guidance . Attachment B provides instructions for inspecting the battery and chargers. and follow the "Performance Checks and Maintenance" section of your user manual, in addition to the information in Attachment B, to ensure your system is available for surgical procedures Medtronic field service personnel will install a new version of the user manual " Medtronic field service personnel will install a component (Mobile View Station \'MVS'') heatsink) as part of the new design of the AC power input circuit. Questions contact Medtronic Technical Support if you have any questions or concerns regarding this matter: USA: (800) 595-9709 or 720-890-3160 Worldwide: +1-720-890-3160 RS.NA VTECHSUPPORT@M EDTRON IC.COM
- Initiated
- 2019-05-01
- Root cause
- Device Design
- Terminated
- 2021-01-19
- Distribution
- Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KOREA KOREA, REPUBLIC OF KUWAIT LEBANON LIBYAN ARAB JAMAHIRIYA LK MEXICO NETHERLANDS NEW ZEALAND NICARAGUA POLAND PORTUGAL PUERTO RICO RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVAKIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM VENEZUELA
- Recall number
- Z-1932-2019
- Recall status
- Terminated
- CDRH record ID
- 173294
- FDA record created
- 2019-07-03
- Recall event number
- 82991