Z-1932-2020: Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance a

Recall record Z-1932-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1932-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
May 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.

Source facts

Action
On April 10, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue. Customers were asked to do the following: - Ensure that Cabling Check and Connectors Check are performed daily, as indicated in the User Routine Checks Program (section 7.4 of the Instructions for Use provided with the system). - In case the strain relief is lost and/or damage to the cable/connector is identified; stop using the system and contact your Philips local representative so that the cable can be replaced. - Please ensure that all staff working with the system are informed of the content of this safety notice and place a copy of it with the Instructions for Use. Philips will be replacing the Stand MVS Trolley Cable (459800943703) in all affected systems by a new Stand MVS Trolley Cable that has an improved strain relief design. This action will start by end of April 2020. You will be contacted by a local Philips representative to schedule an appointment to replace the cable. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377
Initiated
2020-04-10
Root cause
Device Design
Terminated
2021-01-11
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.
Recall number
Z-1932-2020
Recall status
Terminated
CDRH record ID
181039
FDA record created
2020-05-08
Recall event number
85473

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗