Open original FDA record ↗Recall record
- FDA identifier
- Z-1934-2012
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA classification date
- Jul 3, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Aesculap Implant Systems, LLC has initiated a voluntary removal of the S4 Cervical (S4C) Occiput Revision Instrument that is part of the S4C Occipital Instrument Set. The instruments may not thread into the screw properly during a revision surgery due to tolerance issues between the two instruments.
Source facts
- Action
- The firm initiated their recall of this product by issuing an "Important Product Removal Notification" letter to all consignees dated March 7, 2012. The letter described the problem, advised them not to use the product, and instructed consignees to return the instruments to the recalling firm with the attached Distribution Inventory Sheet. Customer Service may be contacted if questions at 1-866-229-3002.
- Initiated
- 2012-03-07
- Root cause
- Component design/selection
- Terminated
- 2012-08-08
- Distribution
- Nationwide distribution, including the states of CA, CO, IL, LA, NV, OH, and TX .
- Recall number
- Z-1934-2012
- Recall status
- Terminated
- CDRH record ID
- 110004
- FDA classified
- 2012-07-03
- Recall event number
- 62240