Open original FDA record ↗Recall record
- FDA identifier
- Z-1934-2020
- Company in FDA record
- Cuattro LLC
- FDA record creation date
- May 8, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential that an image from a previous patient study to show up in a later patient study. This may result in misdiagnosis of the correct patient.
Source facts
- Action
- On October 15, 2019, The firm contacted all affected consignees via phone call. In addition to informing consignees about the recall. The firm provided the following information to consignees: . To prevent a reoccurrence of this problem Cuattro has updated the settings file in the program data folder on your acquisition workstation. This update to the file will prevent any backup images from being downloaded to the acquisition workstation. This was completed for your system on September 6, 2019, service ticket number 70002871
- Initiated
- 2019-10-15
- Root cause
- Device Design
- Terminated
- 2021-04-27
- Distribution
- US: NJ, MI, SC, NC, MI, AR, OUS: None
- Recall number
- Z-1934-2020
- Recall status
- Terminated
- CDRH record ID
- 180899
- FDA record created
- 2020-05-08
- Recall event number
- 85446