Open original FDA record ↗Recall record
- FDA identifier
- Z-1935-2010
- Company in FDA record
- SpineMatrix Inc
- FDA classification date
- Jun 30, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A full design review of the CERSR system was conducted in 7/2009. The audit revealed numerous design verification tests that either did not pass or did not meet the design input requirement. One requirement found that it was possible for the physician to think they were reviewing one patients scan, but were actually reviewing a scan from a different patient. This event did not happen; however,
Source facts
- Action
- On 1/29/2010 all customers were provided a Voluntary Product Recall Notification letter that was sent via U.S. Postal Service Certified Mail and they were contacted by phone to arrange for pick up of the devices. The letter identified the affected product and described the problem. Customers are to check their inventory for the affected product and contact SpineMatrix Customer Service to receive instruction as to how to return the affected product, if the firm did not already contact the customer. Customers must complete the verification form and return it. Questions should be directed to Customer Service.
- Initiated
- 2010-01-29
- Root cause
- Device Design
- Terminated
- 2012-09-19
- Distribution
- Nationwide Distribution -- CA, OH, OR & TX.
- Recall number
- Z-1935-2010
- Recall status
- Terminated
- CDRH record ID
- 91822
- FDA classified
- 2010-06-30
- Recall event number
- 55796