Open original FDA record ↗Recall record
- FDA identifier
- Z-1935-2012
- Company in FDA record
- Greiner Bio-One North America, Inc.
- FDA classification date
- Jul 3, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
: The Quickshield Complete Plus may have a defect (split at the base), which can cause the green safety shield attachment of the Quickshield holder to be loose and unstable.
Source facts
- Action
- greiner bio-one sent a notification letter dated April 2, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the attached Product Disposition Site Confirmation form and fax to 800-726-0052 or email to patech@gbo.com. This form will confirm the affected product has been discarded/destroyed. The firm will replace the product after the completed form is returned to them. Customers with questions should call 800-515-8112. For questions regarding this recall call 704-261-7823.
- Initiated
- 2012-04-02
- Root cause
- Process control
- Terminated
- 2013-02-27
- Distribution
- Nationwide Distribution including IL, MS, OH, PA, TN and VA.
- Recall number
- Z-1935-2012
- Recall status
- Terminated
- CDRH record ID
- 110651
- FDA classified
- 2012-07-03
- Recall event number
- 62447