Skip to content

Public FDA record

Z-1935-2024: LOSPA Patella Component Model/Catalog Number: 01.10.9XX Software Version: N/A Product Description: Patell

Recall record Z-1935-2024 for CORENTEC CO., LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1935-2024
Company in FDA record
CORENTEC CO., LTD
FDA classification date
May 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to unsupported 10 year expiration date.

Source facts

Action
On 03/29/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that on 7/1/2020 the firm changed the labeling of their polyethylene (PE) and patella inserts from a 5 year shelf life to a 10 year shelf life. This change was made without adequate documentation evidence to support the increase in shelf life. Customers are instructed to: 1. Inform individuals within your organization who need to be aware of this device recall. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices. 4. Maintain awareness of this notice internally until all required actions have been completed within your facility. 5. Inform Corentec if any of the subject devices have been distributed to other organizations. a. Please provide Corentec contact details so that Corentec can inform the recipients appropriately. 6. Please complete the attached customer response form (acknowledgement form). It is possible that you no longer have any physical inventory of the impacted product on site. Completing this form will allow us to update our records and will also negate the need for us to send any further communications on this matter. Therefore, please complete the customer response form even if you no longer have any of the subject devices in your physical inventory. 7. Return the completed form to Corentec. For any questions or assistance contact Bobby Pham at 310-488-2886 or via email Bobby.Pham@corentec-US.com.
Initiated
2024-03-29
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of AL, AZ, CA, MI, PA, TX.
Recall number
Z-1935-2024
Recall status
Open, Classified
CDRH record ID
207095
FDA classified
2024-05-31
Recall event number
94380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗