Open original FDA record ↗Recall record
- FDA identifier
- Z-1936-2009
- Company in FDA record
- Abbott Diabetes Care, Inc.
- FDA classification date
- Sep 1, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Low results: Freestyle Lite Blood Glucose Test strip may generate low blood glucose readings when tested with the control solution from lot 0822524 with Freestyle Freedom Lite and Freestyle Lite Blood Glucose Meters.
Source facts
- Action
- An "Important: Medical Device Notification" letter dated June 10, 2009 was issued to consignees. First consignees received notifications via letters to both direct consignees and registered customers of FreeStyle Lite and FreeStyle Freedom Lite blood glucose meter users in the states where the affected test strip lot was distributed. Affected accounts and customers were instructed to discontinue distribution and use of the product. The notification will be posted to the Abbott Diabetes Care website. Direct questions about the recall to Abbott Diabetes Care, Inc. by calling 510-749-5400.
- Initiated
- 2009-06-10
- Root cause
- Process control
- Terminated
- 2011-01-10
- Distribution
- US only (states AL, AZ, CA, FL, GA, IL, IN, MD, NJ, NY, OH, PA, RI, SC, TN, TX and WI).
- Recall number
- Z-1936-2009
- Recall status
- Terminated
- CDRH record ID
- 82937
- FDA classified
- 2009-09-01
- Recall event number
- 52427