Open original FDA record ↗Recall record
- FDA identifier
- Z-1937-2026
- Company in FDA record
- Foundation Medicine, Inc.
- FDA classification date
- Apr 23, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.
Source facts
- Action
- Foundation Medicine, Inc. (FMI) notified consignees of the recall on 03/24/2026 via emailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter. Consignees were instructed to complete and return the provided business reply form, review impacted equipment and verify that the PCT V1.3.1 version of the software has been installed, and if the PCT V1.3.1 has not been installed upon the equipment listed above, to indicate as such on the business reply form below. FMI will follow up to assist with installation. Consignees were also instructed to forward the notification to applicable personnel and if there is any impact that requires the notice to be forwarded to another facility, contact that facility and provide them with the notification letter.
- Initiated
- 2026-03-24
- Root cause
- Software change control
- Distribution
- US distribution to North Carolina and Massachusetts.
- Recall number
- Z-1937-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219291
- FDA classified
- 2026-04-23
- Recall event number
- 98640