Skip to content

Public FDA record

Z-1938-2025: MUSE 5 Systems, Model Numbers 2021274-001, 2023032-005, 2027478-001, Large capacity client-server-based c

Recall record Z-1938-2025 for GE Medical Systems Information Technologies Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1938-2025
Company in FDA record
GE Medical Systems Information Technologies Inc
FDA classification date
Jun 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 4/25/25 was sent to customers. Actions to be taken by Customer/User: MUSE 5 systems have reached their End of Sales/Service Support as of November 1, 2014. Please see attached for the End of Sales/Service Support letter that was previously distributed. GE HealthCare recommends that customers replace or upgrade these systems. If you wish to continue using your MUSE 5 system, you should ensure your IT Department provides ongoing prevention of unauthorized access to the system such as not using a well-known password. If passwords are well-known, please update the credentials according to your organization's password policies or contact GE HealthCare Service or your local Service Representative if necessary. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached Medical Device Notification Acknowledgement Response Form to DCAR.Recall@gehealthcare.com. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2025-04-25
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1938-2025
Recall status
Open, Classified
CDRH record ID
213775
FDA classified
2025-06-10
Recall event number
96830

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗