Skip to content

Public FDA record

Z-1939-2009: The BladderScan BVI 9400 is a portable ultrasound instrument that provides a noninvasive measurement of u

Recall record Z-1939-2009 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1939-2009
Company in FDA record
Verathon, Inc.
FDA classification date
Sep 17, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Three specific issues are being addressed: 1. Calculation and use of an incorrect year in dates after 12/31/2009. 2. Double scans, continuous scans, and double printing. 3. Calibration sensitivity.

Source facts

Action
U.S. customers (including Government) were sent a post card on or about 5/18/2009 and were called about the upgrade. Units are to be upgraded by the customers using an online tool called ScanPoint. Customers may elect to return their units to the firm for upgrading. Canadian customers were phoned by Verathon. For foreign units sold and distributed through one of the recalling firm's related firms, Service Bulletin SB-0014 was sent to the distributors located in The Netherlands, United Kingdom, and Australia. Contact Verathon Customer Care at 1-800-331-2313 if you have any questions.
Initiated
2009-05-18
Root cause
Software design
Terminated
2010-10-07
Distribution
Worldwide - US, Canada, Malaysia, Hong Kong, Singapore, Argentina, New Zealand, Barbados, Dominican Republic, and Costa Rica.
Recall number
Z-1939-2009
Recall status
Terminated
CDRH record ID
83963
FDA classified
2009-09-17
Recall event number
52769

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗