Open original FDA record ↗Recall record
- FDA identifier
- Z-1939-2009
- Company in FDA record
- Verathon, Inc.
- FDA classification date
- Sep 17, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Three specific issues are being addressed: 1. Calculation and use of an incorrect year in dates after 12/31/2009. 2. Double scans, continuous scans, and double printing. 3. Calibration sensitivity.
Source facts
- Action
- U.S. customers (including Government) were sent a post card on or about 5/18/2009 and were called about the upgrade. Units are to be upgraded by the customers using an online tool called ScanPoint. Customers may elect to return their units to the firm for upgrading. Canadian customers were phoned by Verathon. For foreign units sold and distributed through one of the recalling firm's related firms, Service Bulletin SB-0014 was sent to the distributors located in The Netherlands, United Kingdom, and Australia. Contact Verathon Customer Care at 1-800-331-2313 if you have any questions.
- Initiated
- 2009-05-18
- Root cause
- Software design
- Terminated
- 2010-10-07
- Distribution
- Worldwide - US, Canada, Malaysia, Hong Kong, Singapore, Argentina, New Zealand, Barbados, Dominican Republic, and Costa Rica.
- Recall number
- Z-1939-2009
- Recall status
- Terminated
- CDRH record ID
- 83963
- FDA classified
- 2009-09-17
- Recall event number
- 52769