Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the attachment end of the motor and contaminate the surgical wound resulting in soft tissue inflammation.
Source facts
Action
Stryker sent an "Urgent Medical Device Recall Notification" dated April 28, 2016, to all affected customers.
The recall notification included a description of the reason for the recall, affected product, customer responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments at (269) 389-4518.
Initiated
2016-04-28
Root cause
Component design/selection
Terminated
2017-03-24
Distribution
Domestic only:
AL
AR
CA
FL
GA
IL
IN
KS
KY
MD
MI
MO
NC
NM
OH
OK
SC
TN
TX
VA
WA
WI
Recall number
Z-1939-2016
Recall status
Terminated
CDRH record ID
146592
FDA classified
2016-06-09
Recall event number
74170