Samples prepared on the BD SPA II, SN T0076 produced inaccurate sample results due to inaccurate sample dispensed as a result of the cap piercing probe
Source facts
Action
The firm filed an MDR and implemented a field action consisting of a February 2010 notification letter and product replacement of affected probes. A CAPA for the probe design has been implemented. Effectivity is being tracked via mail and follow up phone call.
Initiated
2010-02-19
Root cause
Nonconforming Material/Component
Terminated
2011-05-18
Distribution
Worldwide distribution: USA, Australia, Belgium, China (Hong Kong), Colombia, Jamaica, Korea, Mexico, Singapore and Uruguay.
Recall number
Z-1940-2010
Recall status
Terminated
CDRH record ID
91636
FDA classified
2010-06-30
Recall event number
55654