Open original FDA record ↗Recall record
- FDA identifier
- Z-1940-2014
- Company in FDA record
- Global Cooling Inc
- FDA classification date
- Jul 1, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The external power supply input is overheating at the connection to the freezer.
Source facts
- Action
- Consignees received an Urgent Medical Device Recall letter, dated March 20, 2014, a Recall Response letter, a User Alert, and a label to attach to the freezer power supply input. The letter identified the reason for the recall and how to recognize that the device may fail. Customers are to discontinue use of the device if there are any signs of a loose connection, corrosion, discoloration, or melting; and take the necessary actions identified, including contacting Global Cooling Customer Service at service@stirlingultracold.com. The Recall Return Response was to be completed and returned.
- Initiated
- 2014-03-20
- Root cause
- Use error
- Terminated
- 2015-07-01
- Distribution
- Worldwide Distribution -- USA, including the states of AK, AR, AZ, CA, CO, CT, FL, GA, GU, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Australia, Brazil, Canada, China, India, Indonesia, Japan, Korea, Kuwait, Mexico, Netherlands, Pakistan, Singapore, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
- Recall number
- Z-1940-2014
- Recall status
- Terminated
- CDRH record ID
- 127369
- FDA classified
- 2014-07-01
- Recall event number
- 68333