Open original FDA record ↗Recall record
- FDA identifier
- Z-1941-2010
- Company in FDA record
- Steris Corporation
- FDA classification date
- Jun 30, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
STERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table
Source facts
- Action
- On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.
- Initiated
- 2010-05-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-02-24
- Distribution
- Worldwide distribution: USA, Australia, India, Korea & New Zealand
- Recall number
- Z-1941-2010
- Recall status
- Terminated
- CDRH record ID
- 92033
- FDA classified
- 2010-06-30
- Recall event number
- 55850