Open original FDA record ↗Recall record
- FDA identifier
- Z-1941-2011
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Apr 11, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sensor cables may suffer material degradation when exposed to certain sterilization procedures.
Source facts
- Action
- GE OEC Medical Systems, Inc sent an Urgent Recall Notice dated March 1, 2011, by certified letter on February 24, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised if their Transmitter Sensor Cables or Receiver Sensor Cables have been exposed to any STERRAD sterilizer other than STERRAD 1005, discontinue use of these sterilizers for this product. The Transmitter Sensor Cables should be inspected for signs of material degradation. Further information is available at 800-874-7378.
- Initiated
- 2011-02-24
- Root cause
- Component design/selection
- Terminated
- 2011-09-09
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI and WV and the countries of Austria, Canada, China, Denmark, England, France, Germany, Hong Kong, India, Ireland, Italy, Japan, San Salvador, Slovakia, Spain, and Switzerland.
- Recall number
- Z-1941-2011
- Recall status
- Terminated
- CDRH record ID
- 98340
- FDA classified
- 2011-04-11
- Recall event number
- 58098