Open original FDA record ↗Recall record
- FDA identifier
- Z-1941-2012
- Company in FDA record
- Pride Mobility Products Corp
- FDA classification date
- Jul 3, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.
Source facts
- Action
- Three (3) pieces of information was distributed in the notification. They are: (1) Letter of Correction to the Provider (2) Letter of Correction for the Consumer (provider to issue to consumer) (3) Product Insert for Victory 9 Controller Interface Harness Replacement. The letter of Correction for Victory 9 with Power Sear (SC609PS), dated November 2010, identified the affected product and the reason for the correction. Customers were to record the serial number(s) and contact their provider's office to make an appointment to have the harness replaced.
- Initiated
- 2010-11-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-22
- Distribution
- Worldwide Distribution
- Recall number
- Z-1941-2012
- Recall status
- Terminated
- CDRH record ID
- 95926
- FDA classified
- 2012-07-03
- Recall event number
- 57350