Z-1941-2014: illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed

Recall record Z-1941-2014 for Meridian Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1941-2014
Company in FDA record
Meridian Bioscience Inc
FDA classification date
Jun 27, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.

Source facts

Action
Meridian sent an Urgent Medical Device Recall letter dated April 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return their unit to Meridian Bioscience. Meridian's Technical Support Staff will call them directly to provide them with a return authorization number and instructions for decontaminating, packaging and shipping the unit. For additional information customers were instructed to call 800-343-3858 (North America or 513-271-3700 (Worldwide).
Initiated
2014-04-22
Root cause
Employee error
Terminated
2014-06-30
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.
Recall number
Z-1941-2014
Recall status
Terminated
CDRH record ID
127512
FDA classified
2014-06-27
Recall event number
68396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗