Open original FDA record ↗Recall record
- FDA identifier
- Z-1942-2009
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Sep 16, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The tether attaching to the double swivel cap may partially or completely break at the attachment points, resulting in the potential of the accidental aspiration of the tether into the patient's lungs.
Source facts
- Action
- Teleflex Medical, Inc. notified consignees via overnight mail starting July 31, 2009 requesting acknowledgement of the recall and a request to return all of the affected product. Distributors were instructed to forward the letter to their consignees and retrieve relevant product from those locations. Affected product received by Teleflex Medical will be destroyed or reworked and replacement product or credit will be provided to the consignee. For further information, contact Teleflex Medical at 1-919-433-4986.
- Initiated
- 2009-07-31
- Root cause
- Process control
- Terminated
- 2011-09-15
- Distribution
- U.S. Nationwide, Canada, Mexico, Europe, India, Asia Pacific Rim, South and Latin America. (Including: Malaysia, Hong Kong, Vietnam, Singapore, Thailand, Korea, China, Japan, Philippines, Australia, New Zealand, France, Germany, Ireland, Sweden, Costa Rica, Puerto Rico, Jamaica, Brazil, Argentina, Uruguay and Colombia).
- Recall number
- Z-1942-2009
- Recall status
- Terminated
- CDRH record ID
- 84077
- FDA classified
- 2009-09-16
- Recall event number
- 52823