Open original FDA record ↗Recall record
- FDA identifier
- Z-1942-2014
- Company in FDA record
- Blue Belt Technologies MN
- FDA classification date
- Jun 27, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mislabeled. Packages labeled PFSI-00053, Right Lateral/Left Medial contains PFSI-00054, Right Medial/Left lateral implants. Packages labeled PFSI-00054, Stride Right Medial/Left Lateral contain PFSI-00053, Left Medial/Right Lateral.
Source facts
- Action
- Firm initiated recall on 6/4/2014 by Blue Belt Technologies representatives hand delivering an Urgent Medeical Device Recall letter. The letter was addressed to the attention of Safety Officer/Regulatory Compliance Officer. The letter described the issue of the labeling mix-up, asked for devices to be quarantined, and stated that a Blue Belt Technologies Clinical Specialist will be contacting the consignee or return and replacement. A Receipt Acknowledgement form was asked to be returned by fax or mail.
- Initiated
- 2014-06-04
- Root cause
- Labeling mix-ups
- Terminated
- 2014-08-04
- Distribution
- US Nationwide Distribution in the states of CA, PA, and TX
- Recall number
- Z-1942-2014
- Recall status
- Terminated
- CDRH record ID
- 128024
- FDA classified
- 2014-06-27
- Recall event number
- 68482