Skip to content

Public FDA record

Z-1943-2015: Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used t

Recall record Z-1943-2015 for US Endoscopy Group Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1943-2015
Company in FDA record
US Endoscopy Group Inc
FDA classification date
Jul 1, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation of the components during use.

Source facts

Action
US Endoscopy sent an Urgent Recall Notification Letter dated to their customers. The letter identified the affected product, problem and actions to be taken. US Endoscopy requests that their customers: 1. Remove all unused affected Velocity Biopsy Valves from your inventory. 2. Complete the attached Customer Response Card and return to the attention of Kia Pennington, via fax 1-440-639-4495 or email kpenning@usendoscopy.com. 3. Mark the shipping container with code: RMA xxxxxxxx QIF, and return the product to US Endoscopy via Fed Ex Account #293799547. For questions contact Kia Pennington in Customer Service at 1-800-769-8226.
Initiated
2015-02-17
Root cause
Under Investigation by firm
Terminated
2015-09-21
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, CO, IA, MA, OH, TX & WV and the countries of: Brazil, Costa Rica, Croatia, Czech Republic, Egypt, France, Germany, India, Indonesia, Italy, Latvia, Lithuania, Philippines, Poland, Romania, Russian Federation, Slovenia & South Africa.
Recall number
Z-1943-2015
Recall status
Terminated
CDRH record ID
137329
FDA classified
2015-07-01
Recall event number
71336

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗