Z-1943-2021: inLight Medical LED Pads used with Polychromatic light technology system

Recall record Z-1943-2021 for Inlightened Partners LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1943-2021
Company in FDA record
Inlightened Partners LLC
FDA record creation date
Jun 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.

Source facts

Action
On 6/5/2019, Medical Device Recall Letters were emailed to customers. Customers were instructed to examine the connector on the cord attached to their pad and compare it to photos provided in the Medical Device Recall Letter. The connectors for recalled product contain a v-shaped notch. Customers in possession of a pad with the connector matching the photos were requested to complete an Acknowledgement and Receipt Form. Customers with questions were asked to call Customer Service at 1-800-455-4116.
Initiated
2019-06-04
Root cause
Component change control
Terminated
2021-06-25
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, TX, NC, ID, CO, VA, FL, VT, AK, DC, MA, IL, AZ, MN, IA, TN, WI, WA, OK, MI, OH, GA, NY, IN, PA, NV, NJ, AR, CO , LA, MT, OR, SD, ND, MD, UT, MO, KY, NM, KS, AS, NE, WY, PR, MS and the country of Canada.
Recall number
Z-1943-2021
Recall status
Terminated
CDRH record ID
180100
FDA record created
2021-06-24
Recall event number
85106

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗