Open original FDA record ↗Recall record
- FDA identifier
- Z-1943-2021
- Company in FDA record
- Inlightened Partners LLC
- FDA record creation date
- Jun 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.
Source facts
- Action
- On 6/5/2019, Medical Device Recall Letters were emailed to customers. Customers were instructed to examine the connector on the cord attached to their pad and compare it to photos provided in the Medical Device Recall Letter. The connectors for recalled product contain a v-shaped notch. Customers in possession of a pad with the connector matching the photos were requested to complete an Acknowledgement and Receipt Form. Customers with questions were asked to call Customer Service at 1-800-455-4116.
- Initiated
- 2019-06-04
- Root cause
- Component change control
- Terminated
- 2021-06-25
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, TX, NC, ID, CO, VA, FL, VT, AK, DC, MA, IL, AZ, MN, IA, TN, WI, WA, OK, MI, OH, GA, NY, IN, PA, NV, NJ, AR, CO , LA, MT, OR, SD, ND, MD, UT, MO, KY, NM, KS, AS, NE, WY, PR, MS and the country of Canada.
- Recall number
- Z-1943-2021
- Recall status
- Terminated
- CDRH record ID
- 180100
- FDA record created
- 2021-06-24
- Recall event number
- 85106