Z-1944-2020: iChemVELOCITY Urine Chemistry System-

Recall record Z-1944-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1944-2020
Company in FDA record
Beckman Coulter Inc.
FDA classification date
May 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

Source facts

Action
On 03/25/2020 Beckman Coulter Inc. sent out Urgent Medical Device Alert identifying the problem, the circumstances in which it can occur, the actions that should be taken by the customer/user in order to prevent risks to patients or users, and the actions planned by Beckman Coulter to correct the problem. PRODUCTS: Products installed between July of 2012 and December of 2014 are in scope of this field corrective action and can be identified by the name of the Product (all part numbers) that were installed in the time frames noted. ACTION TO BE TAKEN BY CUSTOMER/USER: If flames and or smoke be detected within the computer the customers/users should; follow laboratorys safety guidelines, contact Beckman Coulter by phone: call 800-526-7694 in the United States and Canada or if outside the United States and Canada, contact Beckman Coulter hotline or the local Beckman Coulter representative, and Follow Standard Operating Procedures (SOP) for Instrument backup testing methodology. This information should be communicated to all laboratory staff and retained as part of lab quality system documentation. Customers are asked to complete the Notification Response Form either electronically or manually depending on the method the notice was received. ACTIONS PLANNED BY BECKMAN COULTER: A Beckman Coulter service provider will proactively schedule a service visit to inspect the PC6 computer and replace the affected cable, if necessary. Additional potentially affected customers will be notified of the issue via Updated Urgent Medical Device Recall letter that will be sent during the week of 10/26/2020 via mail and/or e-mail. Note: Customers that received the initial letter but have any of the additional potentially affected devices will be re-notified of the issue.
Initiated
2020-03-30
Root cause
Nonconforming Material/Component
Terminated
2023-08-31
Distribution
US Distribution: Nationwide Foreign Distribution: Worldwide
Recall number
Z-1944-2020
Recall status
Terminated
CDRH record ID
180946
FDA classified
2020-05-09
Recall event number
85423

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗