Open original FDA record ↗Recall record
- FDA identifier
- Z-1945-2013
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 9, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system.
Source facts
- Action
- The firm, Siemens, sent an "Important safety notice" by letter on May 13, 2013 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to immediately notify the service department to clarify what further measures need to be taken, if after carrying out the regular brake test, the system can only be moved at a reduced speed. If this device has been sold and is therefore no longer in your possession, send this safety notice to the new owner and provide the new owner to Siemens. Update instruction AX012/13/S was issued, and a software update will be released as AX006/13/S and AX007/13/S which will modify the system software so that it will not be possible for information to be overlaid. If you have any questions call +49 (9191) 18 0.
- Initiated
- 2013-05-13
- Root cause
- Software design
- Terminated
- 2014-11-06
- Distribution
- Nationwide distribution: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, LA, MA, MD, MN, MO,MS, NC, NE, NH, NJ, NY, OH, PA, PR, SD, TN, TX, UT, VA, WA, and WI.
- Recall number
- Z-1945-2013
- Recall status
- Terminated
- CDRH record ID
- 119960
- FDA classified
- 2013-08-09
- Recall event number
- 65721