Open original FDA record ↗Recall record
- FDA identifier
- Z-1945-2019
- Company in FDA record
- Bio-Detek, Inc.
- FDA record creation date
- Jul 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some of the electrodes may have been assembled incorrectly, preventing delivery of defibrillation or pacing therapy.
Source facts
- Action
- The firm notified customers of the recall on May 28, 2019, via Urgent Device Correction letter. Customers were advised to take the following actions: - Inform users of the problem. -Remove all stock of part numbers 8900-2105-01 and 8900-2106-01 with lot numbers 1719, 1719A , and 1819. - Complete the response form and return to ZOLL. A ZOLL representative will contact you to coordinate the return and replacement of your electrodes. Customers may direct their questions to the firm's 24/7 technical support numbers 1 (800) 348-9011 or +1(978)421-9460
- Initiated
- 2019-05-28
- Root cause
- Process control
- Terminated
- 2020-06-01
- Distribution
- Distributed to accounts in AZ, CA, CO, CT, FL, LA, MN, NV, OH, PA, TN, UT, and WY. International distribution to Canada, Germany, France, and Italy.
- Recall number
- Z-1945-2019
- Recall status
- Terminated
- CDRH record ID
- 173389
- FDA record created
- 2019-07-03
- Recall event number
- 83040