Open original FDA record ↗Recall record
- FDA identifier
- Z-1945-2024
- Company in FDA record
- Ion Beam Applications S.A.
- FDA classification date
- Jun 3, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Source facts
- Action
- Iba issued Urgent Medical Device Correction Letter on May 10, 2024 via IBA responsible directly to customer site either by hand or by email. Letter states reason for recall, health risk and action to take: ACTION BEING TAKEN BY IBA Monitoring of the deflector, software upgrade. 7.2 CORRECTIVE ACTIONS Software upgrade 7.3 PREVENTIVE ACTIONS Waiting for the solution to be deployed, IBA will distribute an Internal User Notice to IBA operators of impacted sites with instructions to monitor the deflector and stop the irradiation when necessary. 7.4 TIME SCHEDULE FOR THE IMPLEMENTATION OF THE DIFFERENT ACTIONS The Internal User Notice will be provided to IBA operators by May 10, 2024. The corrective action will be implemented at your site by June 2025. Contact: Sonia PINEL Vigilance@iba-group.com
- Initiated
- 2024-05-10
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution in the state of Virginia and the country of India.
- Recall number
- Z-1945-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207610
- FDA classified
- 2024-06-03
- Recall event number
- 94576