Open original FDA record ↗Recall record
- FDA identifier
- Z-1945-2026
- Company in FDA record
- Clinical Innovations, LLC
- FDA classification date
- Apr 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to complaints of device breakage at the traction force gauge to handle joint.
Source facts
- Action
- On 03/10/2026, the firm emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that there is a potential of failure on the handle breaking at the Traction Force Indicator handle joint on the Kiwi Complete Vacuum Delivery System (VAC-6000MT) when angular force is applied to the handle. Customers are instructed to: 1. Identify and quarantine all unused inventory of the affected product. 2. Discontinue distribution of the affected product. 3. Return all unused, affected product to Clinical Innovations (Laborie), regardless of the labeled expiration date. 4. Complete and return the attached Recall Acknowledgement and Receipt Form within 15 business days, even if no inventory remains. Return the forms via email to recalls@Laborie.com. 5. If customers have product to return, a company representative will contact them once Clinical Innovations has receive their Recall Acknowledgement and Receipt Form to issue them an RMA per company policy and arrange for a product return. For questions, call Clinical Innovations (Laborie), at 1-(888)- 268-6222 M-F 8:00 AM- 5:00 PM MT or + (33) 383 22 20 76 M-F 8:00 AM - 5 PM GMT +2 or their Clinical Innovation's (Laborie) service representative.
- Initiated
- 2026-03-09
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MI, MN, MS, NC, NE, NH, NV, NY, OH, PR, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Canada, China, Japan, Czech Republic, and Singapore.
- Recall number
- Z-1945-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219144
- FDA classified
- 2026-04-24
- Recall event number
- 98597