Complaints were received from the field indicating that some Air Pumps were flashing and alarming when connected to the IsoGel AIR" surface upon turning the pump on and the setting it to the IsoGel AIR Mode. Upon further investigation, it was determined that the flashing/alarm being experienced was due to the air pressure being out of the expected range; indicating a blockage within the LAL tubi
Source facts
Action
On 7/12/13, an URGENT MEDICAL DEVICE RECALL NOTIFICATION was mailed to consignees. The notification addressed the affected product and issue, potential hazard, actions needed including faxing (269-488-8691) or emailing (ProductFieldAction@stryker.com) the attached reply form to confirm receipt of notification, and risk mitigation. Please contact the firm at 269-389-8129 if you have questions or concerns regarding this recall. Business hours are M-F 8am-5pm(ET).
Initiated
2013-07-12
Root cause
Nonconforming Material/Component
Terminated
2014-09-11
Distribution
Nationwide Distribution-USA including the states of OH, ME, IL, CO , FL, TX, MI, GA, and VA.
Recall number
Z-1946-2013
Recall status
Terminated
CDRH record ID
118586
FDA classified
2013-08-09
Recall event number
65337