According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Source facts
Action
Newport will contact each affected customer, distributor and end user, to determine the status of the laser s current hardware and whether the power supply must be replaced. If the power supply has the incorrect hardware and must be replaced, Newport will remedy the defect by replacing the unit with a compliant unit.
Initiated
2024-03-27
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide
Recall number
Z-1946-2024
Recall status
Open, Classified
CDRH record ID
208167
FDA classified
2024-06-05
Recall event number
94748
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.