Open original FDA record ↗Recall record
- FDA identifier
- Z-1946-2026
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Apr 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Source facts
- Action
- On March 30, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Risk Managers were instructed to notify affected personnel and locate and quarantine affected product in inventory. A Zimmer Biomet sales representative may remove and return the affected product on your behalf, or the customer may choose to directly return all affected product from their facility. Surgeons were informed that there are no specific patient-monitoring instructions recommended beyond existing follow up schedules.
- Initiated
- 2026-03-30
- Root cause
- Vendor change control
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.
- Recall number
- Z-1946-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219397
- FDA classified
- 2026-04-24
- Recall event number
- 98665