Skip to content

Public FDA record

Z-1947-2009: ACT3020 Advanced Cutting technology (ACT) Twist Drill 3.0mm(D) X 20mm(L). BIOMET 3i 455 Riverside Drive,

Recall record Z-1947-2009 for Biomet 3i, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1947-2009
Company in FDA record
Biomet 3i, LLC
FDA classification date
Sep 1, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some of the packages of ACT3020, drills may actually contain an ACT3220 drill.

Source facts

Action
Consignees were notified of recall by telephone and by faxed an Urgent: Medical Device Recall letter, dated June 30, 2009. The letter stated the problem. The letter also asked customers to check their surgical kits and confirm the drill is placed in the correct space within the kit. Customers are to verify that the shank of the drill is marked as ACT3020 on one side and as 3.0 x 20mm on the opposite side. The letter also stated the clinical implications. Customers are to respond with the attached response fax and then return the drills to BIOMET 3i. Once the returned product is received, replacement drills will be sent to the customers. Questions or concerns should be directed to Kelly Taylor at 561-776-6906.
Initiated
2009-06-29
Root cause
Mixed-up of materials/components
Terminated
2010-01-05
Distribution
Worldwide Distribution -- Korea, Netherlands, Taiwan, United Kingdom, and USA.
Recall number
Z-1947-2009
Recall status
Terminated
CDRH record ID
83447
FDA classified
2009-09-01
Recall event number
52604

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗